6

D

ata Mana

g

e

ment

Data Management

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

6-3

©2006-2013 Physio-Control, Inc.

Figure 6-1

is an example of a CODE SUMMARY report. Press

CODE SUMMARY

to print the

report.

Figure 6-1

CODE SUMMARY Report

Preamble

The preamble contains patient information (name, event identification, patient identification,
location, age, and sex) and device information (date, time, and therapy information) as shown in

Figure 6-1

. The event identification is a unique identifier that the defibrillator automatically enters

in the ID field for each Patient Report. This identifier includes the date and time that the
defibrillator is turned on. The location field allows you to enter up to 25 alpha-numeric characters
to identify where the device was used. You can link the data you enter to other patient
information.

Event/Vital Signs Log

The LIFEPAK 20e defibrillator/monitor documents events and vital signs in chronological order.
Events are operator or device actions that are related to monitoring, pacing, AED therapy, data
transmission, and more.

Table 6-1

shows a complete listing of events that can be found in the

event log.

Vital signs (or active parameters) are entered into the log automatically every 5 minutes (or for
each event) and when alarm limits are exceeded. Refer to

AUTO LOG

in Table 8-1 for more

information.

Preamble

Name:

DAVIDO, GUIDO

CODE SUMMARY™

ID

041495094322

critical event record

Patient ID:

52876004

Power On:

24 April 09 06:03:12

Location:

L483

Device:

100

Age: 45

Sex: M

Site:

ABCD

Total Shocks:

3

Total time paced

00:15:00

Elapsed Time:

00:52:43

35.1 3434 LP20PRB005

Event/Vital Signs Log

Time

Event

HR

SpO2•PR

COMMENTS:

07:15:34

Power On

07:16:34

Initial Rhythm

95

9995

07:20:34

Vital Signs

92

9894

07:22:14

Pacing 1 Started

95

9895

07:24:34

Pacing 2 Set

99

9899

07:25:34

Vital Signs

92

9893

07:26:36

Alarm HR

152

99