LIFEPAK 20e Defibrillator/Monitor Operating Instructions

4-11

©2006-2013 Physio-Control, Inc.

• If you did witness the arrest, you should press

ANALYZE

and proceed directly to analysis. This

will end the CPR period and you will see and hear

ANALYZING NOW, STAND CLEAR

.

Note: The decision to end CPR early is based on your hospital protocol.

• If you did not witness the arrest, you should perform CPR and not press

ANALYZE

. The

INITIAL CPR

countdown timer continues for the duration specified in the

INITIAL CPR TIME

setup option, for example, 90 seconds. When

INITIAL CPR TIME

ends, you will see and hear

PUSH ANALYZE

.

Initial CPR - Analyze First

When the

INITIAL CPR

option is set to

ANALYZE FIRST

, you will be prompted to perform analysis

after the AED is turned on. CPR is prompted after the AED completes the analysis.

If the electrodes are not attached to the patient, you will see and hear

CONNECT ELECTRODES

before you are prompted to perform analysis.

No Shock Advised

When initial CPR time ends, you will see and hear

NO SHOCK ADVISED

followed by

PUSH

ANALYZE

.

After 3 seconds, a countdown timer
continues for the duration specified in
the initial CPR period and you will see
and hear

IF YOU WITNESSED THE

ARREST, PUSH ANALYZE

. This provides

an opportunity to end the initial CPR
early and proceed to analysis.

If the AED detects a nonshockable
rhythm, you will be prompted to

START

CPR.

A countdown timer (min:sec

format) continues for the duration
specified in the

INITIAL CPR TIME

setup

option. (Refer to

Section 8

).