Specifications and Performance Characteristics

A-12

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

CHARGE TIME

AC Operation Only:

Maximum Time from Charge to Shock Ready

Voltage

Manual Mode

AED Mode

100–120V

360 J

7 seconds

200 J

5 seconds

220–240V

360 J

7 seconds

200 J

5 seconds

90V (90% of Nominal -100)

360 J

7 seconds

200 J

5 seconds

198V (90% of Nominal -220) 360 J

7 seconds

200 J

5 seconds

Maximum Time from Initiation of Analysis to Shock Ready (AED only)

Voltage

AED Mode

100–120V

200 J

12 seconds

220–240V

200 J

12 seconds

90V (90% of Nominal -100)

200 J

12 seconds

198V (90% of Nominal -220)

200 J

12 seconds

Maximum Time from Power-on to Shock Ready

Voltage

Manual Mode

AED Mode

90V (90% of Nominal -100)

360 J

11 seconds

200 J

16 seconds

198V (90% of Nominal -220) 360 J

11 seconds

200 J

16 seconds

Battery-only Operation:

Maximum Time from Charge to Shock Ready

Battery

Manual Mode

AED Mode

Fully charged

360 J

7 seconds

200 J

5 seconds

Fully charged battery,
followed with 15 full energy
discharges

360 J

7 seconds

200 J

5 seconds

Maximum Time from Initiation of Analysis to Shock Ready (AED only)

Battery

AED Mode

Fully charged

200 J

12 seconds

Fully charged battery, followed with
15 full energy discharges

200 J

12 seconds

Maximum Time from Power-on to Shock Ready

Battery

Manual Mode

AED Mode

Fully charged battery,
followed with 15 full energy
discharges

360 J

11 seconds

200 J

16 seconds