Noninvasive pacing procedure, Noninvasive pacing procedure -26 – Physio-Control LIFEPAK 20e User Manual

Page 96

Advertising
background image

Therapy

4-26

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

The LIFEPAK 20e defibrillator/monitor has an integrated pulse oximeter that can be used in
conjunction with a noninvasive pacemaker to help confirm capture. To confirm capture, compare
the pulse rate measured by the oximeter to the set pacing rate of the pacemaker.

Noninvasive Pacing Procedure

ECG monitoring during pacing must be performed with the ECG electrodes and patient ECG
cable. Pacing therapy electrodes cannot be used to monitor ECG rhythm and deliver pacing
current at the same time. Be sure to place the therapy electrodes in the proper locations as
described in the pacing procedure. Improper electrode placement may make a difference in the
capture threshold.

To pace, perform the following:

1 Press

ON

.

2 Connect the patient ECG cable, apply ECG electrodes to the ECG cable and patient, and

select Lead I, II, or III. To receive the best monitoring signal, ensure there is adequate space
between the ECG electrodes and the therapy electrodes.

3 Identify the QUIK-COMBO electrode sites on the patient. For pacing, use either the anterior-

lateral or anterior-posterior position (refer to

page 4-3

).

4 Prepare

patient's

skin for electrode application as described in Step 3 of the Defibrillation

Procedure.

5 Apply QUIK-COMBO electrodes to the patient.

6 Connect the therapy electrodes to the therapy cable.

7 Press

PACER

. Confirm the LED illuminates, indicating that the power is on.

Note: If the

REMOVE TEST PLUG

message appears, disconnect test plug and connect

therapy electrodes to QUIK-COMBO therapy cable.

8 Observe the ECG rhythm. Confirm that a triangle sense marker appears near the middle of

each QRS complex. If the sense markers do not appear or are displayed in the wrong location
(for example, on the T-wave)

,

select another lead. (It is normal for the sense marker location

to vary slightly on each QRS complex.)

9 Press

RATE

or rotate the Speed Dial to select the desired pacing rate.

Note: The

RATE

button changes the rate in 10 pulse per minute (ppm) increments; the

Speed Dial changes the rate in 5 ppm increments.

10 Press

CURRENT

or rotate the Speed Dial to increase current until electrical capture occurs.

For each delivered pacing stimulus, the

PACER

indicator flashes off and a positive pace

marker displays on the ECG waveform.

11 Palpate the patient's pulse or check blood pressure and compare the SpO

2

pulse rate with the

set pacing rate to assess for mechanical capture. Consider use of sedation or analgesia if
patient is uncomfortable.

Note: The

CURRENT

button changes the current in 10 mA increments; the Speed Dial

changes the current in 5 mA increments.

Note: To change rate or current during pacing, press

RATE

or

CURRENT

, and then rotate

the Speed Dial.

Note: To interrupt pacing and view the patient's intrinsic rhythm, press and hold the

PAUSE

button. This causes the pacer to pace at 25% of the set rate. Release the

PAUSE

button to resume pacing at the set rate.

12 To stop pacing, reduce current to zero or press

PACER

.

Note: To defibrillate and stop noninvasive pacing, press

ENERGY SELECT

or charge the

defibrillator. Pacing automatically stops. Proceed with defibrillation.

Advertising