Metrohm viva 1.0 Manual User Manual

Page 411

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

399

Identification

Description

ELT

Variable for electrode test:

0 = Electrode test has not been performed

1 = Electrode test has been performed once

2 = Electrode test has been performed twice

3 = Electrode test has been performed three times

FIN

Command status: 1 = Command has ended at
least once
, 0; invalid (variable not available) =
Command has never ended

HCR

Largest current measuring range used:

1 = 224 mA | 2 = 20 mA | 3 = 2 mA | 4 = 200 µA |
5 = 20 µA | 6 = 2 µA | 7 = 200 nA |
8 = 20 nA9 = 2 nA | 10= 200 pA

LCR

Lowest current measuring range used:

10 = 200 pA | 9 = 2 nA | 8 = 20n A | 7 = 200 nA |
6 = 2 µA | 5 = 20 µA | 4 = 200 µA | 3 = 2 mA |
2 = 20 mA | 2 = 224 mA

NMP

Number of measuring points in measuring point list

RPM

Stirring rate in min

-1

STY

Type of stop with which the command was stopped:

1 = normal; 0 = manual or after an error

Result variables if called by CALL VA
The following result variables are generated in the method run when a
CVS command is called by CALL VA and can be used in formulas:

Result variables per substance, variation and replication

Designation 'RS.'Command name'.'Substance
name'.VAR{x}.REP{y}.Variable identification'

Identification

Description

SUBST

Substance name

POT

Peak potential

AREA

Peak area

HGT

Peak height