5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

454

■■■■■■■■

viva 1.0

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate.

Input range

0 - 15

Default value

3

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.5.2.5.6.3

MEAS TMF - Measuring parameters

Tab: Method

MEAS TMF Properties... Measuring parameters

Command name

Name of the command.

Entry

25 characters

Measurement
Two types of measurement can be selected from, Measurement with
drift control
or Measurement without drift control.

Selection

Measurement with drift control | Measure-
ment without drift control

Default value

Measurement with drift control

Measurement with drift control
The measurement is carried out drift-controlled if this option is
selected. The measurement is canceled as soon as the defined Signal
drift
or the Stop measured value is reached or the defined Max.
waiting time
has expired. The corresponding parameters are not visi-
ble if this option is disabled.