9.2

Troubleshooting

9.2.1

894 Professional CVS

Problem

Cause

Remedy

An electrode cable
has broken.

The cable has accidentally
been jammed in the mea-
suring head cover or a
cable contact has been
kinked.

Replace the electrode cable set (see Manual
894 Professional CVS
).

Conditioning takes a
long time.

The reference electrode is
not yet equilibrated.

After maintenance of the reference electrode,
wait for at least 20 min before starting a mea-
surement.