Metrohm viva 1.0 Manual User Manual

Page 433

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

421

Input range

0 - 999

Default value

20

Action

Definition of the action that is to be executed if this stop criterion is fulfil-
led.

Selection

Cancel command | Cancel determination | Can-
cel determination and series

Cancel command
The running command CALL COND is canceled, afterwards the next
command is started in the track.
Cancel determination
The ongoing determination is canceled, after which the next determi-
nation in the series is started (any exit track that may be present is exe-
cuted, any error track that may be present is not executed).
Cancel determination and series
The ongoing determination is canceled, the next determination in the
series will no longer be started (any exit track that may be present is
executed, any error track that may be present is not executed).

Run time

on | off (Default value: off)

If this check box is activated, then the defined action is triggered when
the specified run time has been reached.

Run time

Run time after which the call of the VA track is to be canceled.

Input range

0.0 - 999.9 min

Default value

0.0 min

Action

Definition of the action that is to be executed if this stop criterion is fulfil-
led.

Selection

Cancel command | Cancel determination | Can-
cel determination and series

Cancel command
The running command CALL COND is canceled, afterwards the next
command is started in the track.