background image

7.6 Avantes Spectrometer

■■■■■■■■■■■■■■■■■■■■■■

772

■■■■■■■■

viva 1.0

Analog output

Analog output

on | off (Default value: on)

If this check box is activated, then the measuring signal is output at the
analog output of the spectrometer. If the connected spectrometer is not
equipped with an analog output, then this setting will be ignored.

Sensitivity

Input range

1 - 214,783,647 mAU/mV

Default value

100 mAU/mV

7.6.4

Properties - Detector

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Detector

The Detector tab shows information on the detector that is built into the
spectrometer.

Detector type

Shows the detector type.

Minimum wavelength

Shows the minimum wavelength of the measurable range in nm.

Maximum wavelength

Shows the maximum wavelength of the measurable range in nm.

Number of pixels

Shows the resolution with which the spectrum is recorded.

7.6.5

Properties - Calibration

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Calibration

The Calibration tab shows the data of the wavelength calibration. A
wavelength defined by four wavelength coefficients is allocated to each
pixel on the detector. This wavelength calibration is carried out with the
CAL Spec command. In this process, the calibration data is calculated
automatically and displayed on this tab.

Date

Shows the time and date of the wavelength calibration.