3 sample changer - moving, Moving – Metrohm viva 1.0 Manual User Manual

Page 819

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Manual control

viva 1.0

■■■■■■■■

807

Number of positions

Shows the number of positions on the attached rack. If no rack is
attached, -------- is displayed.

[Initialize rack]

The attached rack is initialized.

NOTE

When the rack is initialized, the following actions are carried out:

Rack rotates to the position for reading out the rack code.

Rack data on the rack code is transferred to the sample changer.

Lifts are moved upwards to 0 mm.

The robotic arm is moved back.

8.4.5.3

Sample changer - Moving

Program part: Manual control

Moving to the desired rack position, lift position or robotic arm position
can be triggered manually here. In addition, the shift rate and lift rate for
the manual changer control can be set here.

Rack position
Set the rack position.

Current position

Shows the current rack position.

Target position

Selection or entry of the rack position to be approached.

Input range

1 - n (depending on rack)

Default value

1

Selection

Special beaker 1 - 16

Shift rate

Shift rate for manual control of the sample changer.

Input range

5 - 20 °/s

Default value

20 °/s

[Start]

Starts moving to the target position. After the start, the button changes to
[Stop], the two lower buttons are shown as inactive (gray) and Move... is
shown as status message instead of Ready.