7 rack, 1 rack - overview, 2 rack - properties – Metrohm viva 1.0 Manual User Manual

Page 528: Rack

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

516

■■■■■■■■

viva 1.0

5.5.2.8.7

RACK

5.5.2.8.7.1

RACK - Overview

Dialog window: Method

RACK Properties... RACK - 'Command name'

Command for the initialization of the sample rack in place. The rack
that is in place, the lift and the robotic arm (if present) are reset, the rack
code is read out and the respective rack data is transferred to the sample
changer.

Instruments
This command can be executed with the following instruments:

Sample changer: 814, 815, 858, 919

Appearance
The command has the following appearance:

Parameters
The parameters for the RACK command are configured in the following
dialog window:

RACK - Properties

Command variables
The following command variables are generated by the RACK command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

FIN

Command status; 1 = Command has ended at
least once
, 0, invalid (variable not available) =
Command has never ended

5.5.2.8.7.2

RACK - Properties

Dialog window: Method

RACK Properties... RACK - 'Command name'

Command name

Name of the command.

Entry

25 characters