5 Method

viva 1.0

■■■■■■■■

567

5.6.5.4

MLAT calibration curve properties

Dialog window: Method

Evaluation - Calibration Calibration

curves

[Edit] Properties... Properties calibration

The calibration curve parameters for the substance selected in the table
can be edited in the Properties calibration dialog window.

Measuring command

Shows the selected measuring command.

Substance

Shows the selected substance.

Evaluation quantity

Selection of the size according to which the substance is to be evaluated.

Selection

Area | Height

Default value

Area

Weighting

on | off (Default value: on)

If this check box is activated, then the points on the calibration curve will
be weighted in accordance with the formula 1/evaluation quantity

².

Curve type

Only the Linear regression curve type is permitted for the MLAT cali-
bration method. At least two calibration points of different concentrations
are required for determining the line.

Selection

Linear regression | Quadratic regression | Non-
linear regression
| Linear interpolation

Default value

Linear regression

Curve type

C

ali

br

ati

o
n

fu

nc

ti
o
n