Metrohm viva 1.0 Manual User Manual

Page 468

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

456

■■■■■■■■

viva 1.0

Measuring interval

Time interval for entering a measured value in the measuring point list.

Input range

0.25 - 999,999.00 s (Increment: 0.25)

Default value

2.00 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Input range

0 - 4,000 mAU

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:

Measuring time

Duration of the absorbance measurement.

Input range

0 - 999,999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Input range

0.25 - 999,999.00 s (Increment: 0.25)

Default value

2.00 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Input range

0 - 4,000 mAU

Selection

off

Default value

off

off
No stop.

Advertising