5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

418

■■■■■■■■

viva 1.0

Condition

on | off (Default value: off)

Expression for the condition(s) that can be created or edited with the for-

mula editor after pressing on

or by double-clicking in the text field

(see Chapter 2.3, page 13).

If this check box is activated, then the call is carried out only if the evalua-
tion of the condition yields the result 1 (= true).

Entry

1,000 characters

Default value

'empty'

5.5.2.4.6

CALL COND

5.5.2.4.6.1

CALL COND - Overview

Dialog window: Method

CALL COND Properties... CALL COND - 'Command

name'

Command for the repeated calling of a VA track which continues to
be executed until one of the stop criteria is fulfilled. This command is used
chiefly for the conditioning of electrodes.

Appearance
The command has the following appearance:

Parameters
The parameters for the CALL COND command are configured in the fol-
lowing dialog window:

CALL COND

Command variables
The following command variables are generated by the CALL COND
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started