5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

404

■■■■■■■■

viva 1.0

RDE
Standard electrode of the RDE type, which cannot be deleted.
'Electrode name'
Electrode of the RDE type from the electrode table.

Sensor type

Shows the type of working electrode (RDE).

Reference electrode

Selection of an electrode of the Reference electrode type from the elec-
trodes available in the electrode table.

Selection

Reference electrode | 'Electrode name'

Default value

Reference electrode

Reference electrode
Standard electrode of the Reference electrode type, which cannot
be deleted.
'Electrode name'
Electrode of the Reference electrode type from the electrode table.

Auxiliary electrode

Selection of an electrode of the Auxiliary electrode type from the elec-
trodes available in the electrode table.

Selection

Auxiliary electrode | 'Electrode name'

Default value

Auxiliary electrode

Auxiliary electrode
Standard electrode of the Auxiliary electrode type, which cannot be
deleted.
'Electrode name'
Electrode of the Auxiliary electrode type from the electrode table.

Electrode test

The electrode test for the reference and auxiliary electrodes can be activa-
ted and deactivated only in the START command (see "Electrode test for
VA measuring commands", page 384)
.

Stirrer

Stirring rate

Number of revolutions per minute of the stirrer.

Input range

0 - 3,000 min

⁻¹ (Increment: 100)

Default value

2,500 min

⁻¹