Metrohm viva 1.0 Manual User Manual

Page 515

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

503

Number

Specification of the number of the rack position. This parameter is visible
only when Move = Rack position or Special beaker. The position can,
however, also be entered as a formula. The formula editor is opened with
a right click on the input field.

Move = Rack position
Input range

1 - 999

Default value

1

Move = Special beaker
Input range

1 - 16

Default value

1

Move angle

This parameter specifies the angle by which the sample rack is to be rota-
ted relative to its momentary position. It is visible only when Move = Rel-
ative angle
.

814, 815, 858
Input range

-359.90 - 359.90 °

Default value

5.00 °

Beaker test

Selection of the action to be carried out when the beaker sensor defined
for the rack detects no beaker at the target position. If the robotic arm is
selected as beaker sensor, then a Swing Head must be installed. A work
position that is also defined in the rack data must be defined in such a
way that a contact exists between beaker sensor and beaker. This work
position will be automatically moved to for beaker recognition after the
command MOVE. This section is not visible when Move = Relative
angle
.

Selection

Display message | Cancel determination | Can-
cel determination and series

Default value

Display message

Display message
A message is displayed, the determination is canceled and the next
determination of the series is started after confirmation of the mes-
sage. If an exit track is present, it will be called. No error track will be
called.