5 Method

viva 1.0

■■■■■■■■

491

Port

The port at which the function is to be carried out must be specified for
every Liquid Handling command.

Selection

1 | 2 | 3 | 4

1
Default value for dosing and aspiration
2
Default value for filling, exchange position and port change
4
Default value for expulsion at stop, expulsion at end volume and
compensating

Volume

Volume to be transported. This parameter is available only for the Dosing
or Aspirate functions.

Dosing

Input range

0.00000 - 99,999.9 mL

Default value

10.0000 mL

Aspirate

Input range

0.00000 - 50.0 mL

Default value

10.0000 mL

Rate

Speed with which the functions Dosing, Fill, Aspirate, Eject to stop,
Eject to end volume and Exchange position are carried out. The max-