Metrohm viva 1.0 Manual User Manual

Page 443

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

431

Identification

Description

BP{#}.TIM

Time for the break point # (1 - 9) in s

BP{#}.TRN

Transmission for the break point x (1 - 9) in %

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

DBL

Total duration for the processing of the command in
s

EME

End measured value (measured value after processing
of the command) in the unit of the measured value

ETE

1)

End temperature (temperature following processing
of the command) in °C

FIN

Command status; 1 = Command has ended at
least once
, 0, invalid (variable not available) =
Command has never ended

FP{#}.[MEA]

Measured value for the fixed endpoint # (1 - 9) in the
unit of the measured value

FP{#}.CNT

Intensity for the fixed endpoint x (1 - 9) in counts

FP{#}.TEM

1)

Temperature for the fixed endpoint # (1 - 9) in °C

FP{#}.TIM

Time for fixed endpoint # (1 - 9) in s

FP{#}.TRN

Transmission for the fixed endpoint x (1 - 9) in %

IME

Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value

ITE

1)

Initial temperature (temperature before start condi-
tions are processed) in °C

LP.CAx

Calculated value × (1 - 3) for the last measuring point
on the measuring point list

LP.CNT

Transmission of the last measuring point of the mea-
suring point list in %

LP.EXx

External value × (1 - 3) for the last measuring point of
the measuring point list

LP.MEA

Measured value for the last measuring point of the
measuring point list in the unit of the measured value

LP.TEM

1)

Temperature for the last measuring point of the mea-
suring point list in °C