7.3 858 Professional Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

740

■■■■■■■■

viva 1.0

FPGA version

Shows the FPGA version of the device.

Set to work

Shows the date on which the device was automatically added to the
device table.

Remarks

Remarks about the device.

Entry

1,000 characters

Default value

'empty'

7.3.3

Properties - Tower

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Tower

NOTE

With the 858 Professional Sample Processor and the 919 IC Autosam-
pler plus, the properties of the tower, including the Swing Head
attached to it, are displayed on the Tower tab. In the case of the 814
USB Sample Processor and the 815 Robotic USB Sample Processor XL,
the model versions with one tower have one tab, Tower 1; model ver-
sions with two towers have two tabs, Tower 1 and Tower 2.

Tower parameters

Max. stroke path

Entry of the lowest permitted lift position. A lift height of 0 mm corre-
sponds to the "home position", i.e. the lift is moved all the way to the top.

Input range

0 - 235 mm

Default value

127 mm

Min. beaker radius

Definition of the minimum radius which the beakers used on the rack
must have. If the lift is moved to the work position, the beaker radii
defined in the rack table (see Chapter 6.11.2, page 719) for general sam-
ple positions (see Chapter 6.11.3, page 722) and special beakers (see
Chapter 6.11.3.2, page 722)
are compared with the
Min. beaker
radius
. If this minimum beaker radius is not achieved, the run is stopped
and an error message is output. No check will be made if off is enabled.

Input range

1.0 - 100.0 mm