Metrohm viva 1.0 Manual User Manual

Page 482

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

470

■■■■■■■■

viva 1.0

Parameters for the Avantes spectrometer

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.

Input range

-1E+12 - 1E+12 (max. 10 digits) mAU

Default value

0 mAU

Integration time

Integration time for absorbance measurement.

Input range

0.01 - 600,000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are acquired and averaged.

Input range

1 - 10,000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.

Input range

0 - 100 pixels

Default value

0 pixels

Parameters for the 089 Photometer

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.

Input range

-4,000 - 4,000 mAU

Default value

0 mAU

Advertising