7.8 RS-232 device

■■■■■■■■■■■■■■■■■■■■■■

786

■■■■■■■■

viva 1.0

Delay time

Waiting time until the next character is sent.

Input range

0 - 9,999 ms

Default value

0 ms

[Connect]

Establishes the connection to the RS-232 device. The Establish connec-
tion
dialog window appears (see Chapter 7.8.5, page 788).

[Disconnect]

Disconnects the connection to the RS-232 device (only then can the
device be deleted from the configuration).

7.8.4

Properties - GLP

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

GLP

Tab: Configuration

Dosing units [Edit] Properties... Dosing unit - 'Na-

me'

GLP

Tab: Configuration

Solutions [Edit] Properties... Solution - 'Name' GLP

GLP test date

Date of the last GLP test. This date can be selected by clicking on

in

the Select date dialog window (see Chapter 2.4.1, page 67).

Comment on GLP test

Comment regarding the GLP test.

Entry

1,000 characters

Monitoring of GLP validity
on | off (Default value: off)

If this check box is activated, then the time interval for the GLP test is
monitored.

Monitoring cannot be switched on until after a date has been entered in
the "GLP test date" field.

GLP test interval

Time interval to next GLP test. If a value is entered here, then the date in
the Next GLP test field will be adjusted automatically.

Only editable if monitoring is switched on.
Input range

1 - 999 days

Default value

999 days