5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

494

■■■■■■■■

viva 1.0

Selection

846 Dosing Interface | 814 USB Sample Pro-
cessor
| 815 Robotic USB Sample Processor
XL
| 858.0010 Professional Sample Processor |
858.0020 Professional Sample Processor |
858.0030 Professional Sample Processor |
919.0020 IC Autosampler plus | 894 Profes-
sional CVS
| 089 Photometer

Default value

894 Professional CVS

Dosing device

Dosing device

Selection of the dosing device to be prepared. All the dosing device con-
nections which are possible with the selected device type are always dis-
played.

846, 894
Selection

1 | 2 | 3 | 4

Default value

1

814, 815, 858, 919
Selection

1 | 2 | 3

Default value

1

089
Selection

1

Name of dosing unit

Selection of a dosing unit from the dosing units available in the table of
the dosing units.

Selection

'Name of the dosing unit' | not defined

Default value

not defined

'Name of the dosing unit'
Alphabetically organized selection of the names of the dosing units
from the table of dosing units in the configuration. A check is made
prior to the start of a method to determine whether the selected dos-
ing unit is identical with the name of the dosing unit at the selected
MSB connector.
not defined
No check is made prior to the start of a method to determine whether
the selected dosing unit is identical with the name of the dosing unit at
the selected MSB connector. This has the advantage that a method can
be utilized by users who use other dosing units.