5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

438

■■■■■■■■

viva 1.0

Temperature
This section is displayed only when Device type = 089 Photometer.

Temperature

Measuring temperature which can be entered manually. The temperature
is measured continuously when Temperature measurement = contin-
uous
(tab General/Hardware).

Input range

20.0 - 60 °C

Default value

25.0 °C

5.5.2.5.4.4

MEAS Opt - Evaluations

Tab: Method

MEAS Opt Properties... MEAS Opt - 'Command name' Eval-

uations

Command name

Name of the command.

Entry

25 characters

The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.

Fixed endpoint evaluation

Fixed endpoint evaluation

on | off (Default value: off)

If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value or Time) for the fixed endpoint from the measuring
point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Opens the fixed endpoint evaluation # dialog window to enter a new
fixed endpoint (see Chapter 5.5.2.5.4.6, page 442).

[Properties]

Opens the fixed endpoint evaluation # dialog window to edit the fixed
endpoint selected in the table (see Chapter 5.5.2.5.4.6, page 442).