5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

426

■■■■■■■■

viva 1.0

5.5.2.5.3

MEAS Spec

5.5.2.5.3.1

MEAS Spec - Overview

Dialog window: Method

MEAS Spec Properties... MEAS Spec - 'Command

name'

Command for the Measurement of a sample spectrum.

Instruments
This command can be executed with the following instrument:

Spectrometer: Avantes

Appearance
The command has the following appearance:

Parameters
The parameters for the MEAS Spec command are set on the following
tabs:

General/Hardware
Parameters for instrument.

Measuring parameters
Parameters for setting the measurement procedure.

Evaluations
Definition of evaluation results for the sample spectrum.

Command variables
The following command variables are generated by the MEAS Spec com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':

Identifica-
tion

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

DBL

Total duration for the processing of the command in s

FIN

Command status; 1 = Command has ended at
least once
, 0, invalid (variable not available) =
Command has never ended