Metrohm viva 1.0 Manual User Manual
Page 582
5.6 Evaluation subwindow
■■■■■■■■■■■■■■■■■■■■■■
570
■■■■■■■■
viva 1.0
Input range
0.01 - 1.10
Default value
1.10
Evaluation parameters for sample
Curve type
Curve type for the evaluation of the calibration curve. Determines the type
according to which the calibration curve is to be fitted. A minimum num-
ber p of different effective addition volumes must exist for the sample
solution for the various curve types so that it is possible to fit this calibra-
tion curve. The minimum number p corresponds to the number of terms
in the calibration function.
Selection
Linear regression | Quadratic regression | Non-
linear regression | Linear interpolation
Default value
Linear regression
Selected curve
type
Cali-
bra-
tion
func
tion
p
Linear regression
y =
k(0)
+
k(1)
⋅ x
≥2
Quadratic
regression
y =
k(0)
+
k(1)
⋅ x +
k(2)
⋅ x
2
≥3
Nonlinear
regression
y =
k(0)
+
k(1)
⋅ x +
k(4)
⋅ x
4
≥3