5.6 Evaluation subwindow

■■■■■■■■■■■■■■■■■■■■■■

570

■■■■■■■■

viva 1.0

Input range

0.01 - 1.10

Default value

1.10

Evaluation parameters for sample

Curve type

Curve type for the evaluation of the calibration curve. Determines the type
according to which the calibration curve is to be fitted. A minimum num-
ber p of different effective addition volumes must exist for the sample
solution for the various curve types so that it is possible to fit this calibra-
tion curve. The minimum number p corresponds to the number of terms
in the calibration function.

Selection

Linear regression | Quadratic regression | Non-
linear regression
| Linear interpolation

Default value

Linear regression

Selected curve
type

Cali-

bra-

tion

func

tion

p

Linear regression

y =

k(0)

+

k(1)

x

≥2

Quadratic
regression

y =

k(0)

+

k(1)

x +

k(2)

x

2

≥3

Nonlinear
regression

y =

k(0)

+

k(1)

x +

k(4)

x

4

≥3