AMETEK XG Family Programmable DC User Manual

Page 106

Advertising
background image



Analog

Programming (APG) and Isolated Analog Programming (ISOL)

4-6

M370430-01 Rev E

Pin

Reference

Function

J1.13

VOL_MON

1

Voltage Monitor. The scale of this output is set by the APG voltage
input range selected (see Important note below).

J1.14

CUR_MON

1

Current Monitor. The scale of this output is set by the APG current
input range selected (see Important note below).

J1.15

EXT_CC_CV

External CC/CV. Indicates the state of the operate mode. When in
CV mode, logic high is output and when in CC mode, logic low is
output. Logic is TTL/CMOS compatible.

J1.16

CSH

Current Share. Used to hook up units for current sharing. Output
from master unit. Slave units should be left open.

J1.17

REF_I

Current Reference Input. Slave input from current share output on
the master unit.

J1.18

NC

1. Important: Non-isolated monitor lines are only valid in local control, digital remote and
non-isolated analog programming modes for the respective voltage and current control. If isolated
analog programming is used, the isolated monitor lines should be used as the non-isolated
monitoring lines will not accurately reflect the output state of the power supply.