Calibrating the input voltage apg signal – AMETEK XG Family Programmable DC User Manual

Page 286

Advertising
background image



Calibration

and Troubleshooting

7-22

M370430-01 Rev E

Calibrating the Input Voltage APG Signal

The input voltage APG calibration is necessary to ensure accurate
measurements when using the :MEAS : APR? and :MEAS : APR: ISOL?
SCPI command queries to monitor the APG input signal.

Gain Calibration

To calibrate the VAP APG input gain:

1.

Connect a voltmeter and a DC voltage source to the
non-isolated voltage APG input lines (J1.9 and J1.7).

2.

Turn the power supply On.

3.

Set the Voltage APG level to 4 V.

Send the SCPI command:

SYST:COMM:APR:LEV

4

4.

Set the input source as Voltage analog APG. Send the SCPI

command: SYST:REM:SOUR:VOLT AVOL

5.

Set the voltage source attached to the non-isolated voltage

APG input to 90% of the nominal voltage level, in this case
3.6 V.

6.

Turn the output power on.

7.

The power supply starts up in Constant Voltage mode.

8.

Read the input voltage value from the voltmeter connected in

step 1.

9.

Type the SCPI input Voltage calibration command with

voltage noted in step 8 as the parameter “data”.

For example, you would type :

CAL:

INP

:

ANAL

3.53

if you were

reading 3.53 volts from the voltmeter.

Offset Calibration

Offset calibration of the power supply provides the best accuracy in
low-range input voltage.

To perform offset calibration:

1.

Set the voltage source attached to the non-isolated Voltage
APG input to 10% of the nominal voltage level, in this case
0.400 V.

2.

Turn on the output power.

3.

Read the input voltage value from the voltmeter connected in

step 1 of the gain calibration.

4.

Type SCPI input Voltage calibration command with voltage

noted in step 3 as the parameter “data”.