Isolated current monitoring calibration – AMETEK XG Family Programmable DC User Manual

Page 281

Advertising
background image



Calibration and

Troubleshooting

M370430-01 Rev E

7-17

7

Isolated Current Monitoring Calibration

Important:

For maximum accuracy at a specific APG level, follow the

calibration procedure again but use the intended operating voltage level in steps
3 and 8.

To calibrate the isolated current monitoring:

1.

Short the main output with a shunt. Connect a multimeter to
measure the voltage across the shunt.

2.

Connect the voltmeter to the Isolated monitoring output lines

(J3.10 and J3.6).

3.

Set the current APG level to 4 V. The SCPI command is:

SYST:COMM:APR:LEV:CURR 4

4.

Set the input source to local for voltage analog programming.

the SCPI command is:

SYST:REM:SOUR:CURR LOC

5.

Set the main output voltage to 100% by sending the

command:

SOUR:VOLT <ModelVolt>

or use the rotary Adjust/Enter control.

6.

Turn the main output to On by pressing the OUTPUT

ENABLE Main button on the front panel or send the
command:

:OUTP ON

The OUTPUT ENABLE Main button will illuminate.

7.

Enter the interactive calibration mode by sending the SCPI

command:

CAL:MON:CURR:ISOL

8.

Press the terminal keys - or + to reduce or increase the

monitoring value. Adjust the monitoring value displayed on
the voltmeter that was set up in Step 1 to read as close as
possible to 4.000 V.

9.

Press any key other than the + or - keys to exit from

interactive calibration mode.