Analog Programming

(APG) and Isolated Analog Programming (ISOL)

M370430-01 Rev E

4-11

4

Query for Analog Voltage Input Level

To query for analog voltage input level from non-isolated input:

1.

Turn the 9-position mode control to the VAP position or press
the rotary Adjust/Enter control if the mode control is already
at the VAP position.

is displayed on the output voltage display.

2.

If the mode is currently

then select any other APG

mode. If this step is done, be sure to restore the APG setting to

when the operation is complete. Otherwise, do not

rotate the rotary Adjust/Enter control to keep the current APG
mode.

3.

Press the rotary Adjust/Enter control to select the APG mode.

The output voltage display should read

and the output current

display will display APG level.

4.

Press the rotary Adjust/Enter control to return to normal

mode.

5.

If the APG mode was modified in step 2, execute the

procedure to deactivate the APG voltage mode.

Quick Tip Remote
operation uses these
SCPI commands. See
page 5-1.

The SCPI commands for these instructions are:

[:]SYSTem[<channel>]:REMote:SOURce[:VOLTage] <?|LOCal
|AVOLtage|IAVoltage|ARESistive|IAResistive>

[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel[:VOLTage]
[:ISOlated]

[:]MEASure[<channel>][:SCALar]:APRogram[:VOLTage][:DC]?