Storing calibration data, Step 3: recalibrate gain – AMETEK XG Family Programmable DC User Manual

Page 269

Advertising
background image



Calibration and

Troubleshooting

M370430-01 Rev E

7-5

7

Step 3: Recalibrate Gain

M

e

a

s

u

re

d

V

a

lu

e

Programmed Value

Programmed Value

Figure 7-4

Calibration: Step 3 Recalibrate Gain

Repeat Step 1 for best results.

Storing Calibration Data

Calibration data will not be stored until a CAL:STOR command is
issued. Without calling this command, the new calibration will be lost
after a power cycle.

If desired, the unit may be returned to its calibration state when it
shipped from the factory. To do this, the CAL:REST command may be
issued.

Important:

sThe CAL:REST command will cause the nonvolatile memory of

the supply to be overwritten with the factory calibration state.

Command

Description

[:]CALibration:STORe

Saves calibration data to nonvolatile memory

[:]CALibration:RESTore Resets the calibration to the Factory state