Remote Operation

M370430-01 Rev E

5-63

5

Setting Dwell Time

The dwell time is the amount of time that is delayed between each
command during the execution of an Auto Sequence program. The
dwell time can be from 0 to 180 seconds and can be changed during the
program execution. The dwell time has a minimum step size of 1
second.

Command:

[:]PROGram[<channel>][:STEP]:DWELl {?|<dwell time> }

Where:

<dwell time > is an integer value from 0 to 180.

For Example:

:PROG:DWEL 12 Results in 12 seconds of wait time after each
command is executed.

Storing an Auto Sequence Program

Loading a program into memory is done by using the

[: ] PROGram

[<channel>] [: RECord] : STARt

command. After the start

command has been issued, the XG will record all subsequent commands
into the auto sequence program buffer. The commands will be recorded
into the buffer until the

: PROGram [<channel>] [: RECord] : STOP

command is issued.

Commands:

[:]PROGram[<channel>][:RECord]:STARt

[:]PROGram[<channel>][:RECord]:STOP

Important:

If the start recording command is executed then any previously

stored Auto Sequence program is deleted.

For example:

:PROG:STAR

:VOLT 12

:OUTP ON

:PROG:DWEL 60

:OUTP OFF

:OUTP ON

:OUTP OFF

:PROG:DWEL 0

:PROG:STOP

This program will send a 12 V square wave with 120 seconds 50% duty
cycle. When the program is finished, the dwell time is restored to 0
seconds.