Defining the OVP Set Point

To define the OVP set point:

1.

Turn the power supply On.

Ensure the voltage is lower than the desired set point.

2.

Set the output to the desired voltage.

OVP can be set without setting desired output voltage first.

3.

Turn the 9-position mode control to the PRT position.

PrO

OVP

is displayed.

4.

Press the rotary Adjust/Enter control.

5.

Turn the rotary Adjust/Enter control to set the upper limit
protection. See “Coarse and Fine Adjustment Modes” on
page 3-3.

Important:

The OVP range is variable from VSET +5% to VMAX +25%.

6.

Once the desired value has been set, press the rotary

Adjust/Enter control to commit the setting.

The SCPI command (s) for these instructions are:

[[:]SOURce[<channel>]]:VOLTage:PROTection[:OVERvoltage]
[:LEVel]

If the output is between OVP

set

and -5%, an OVP warning message

is possible when in APG mode.