Offset Calibration

Offset calibration of the power supply provides the best accuracy
in low-range input current.

To perform offset calibration:

1.

Set the voltage source attached to the non-isolated current
APG input to 10% of the nominal current level, in this case
0.400 V.

2.

Turn the output power on.

3.

Read the input voltage value from the voltmeter connected in

step 2 of the gain calibration.

4.

Type SCPI input current calibration command with voltage

noted in step 3 as the parameter “data”.

For example, you would type

:CAL:INP:ANAL:CURR 0.43

.

After performing offset calibration, it is highly recommended that
you repeat gain calibration.

SCPI Command

[:]CALibration[<channel>]:INPut:ANALog:CURRent{<voltme
ter reading in volts>}