Calibration

and Troubleshooting

7-10

M370430-01 Rev E

Non-isolated Analog Programming Calibration

Prior to this, the main output must be calibrated first.

Important:

In calibration commands, when + or - keys are expected, any other

key will exit from the calibration mode.

Non-isolated Voltage Monitoring Calibration

Important:

For maximum accuracy at a specific APG level, follow the

calibration procedure again but use the intended operating voltage level in steps
3 and 8.

To calibrate the non-isolated voltage monitoring:

1.

Disconnect the load and connect the voltmeter to the output.

2.

Connect the voltmeter to the non-isolated voltage monitoring

output lines (J1.13 - J1.7).

3.

Set the Voltage APG level to 4 V. Send the SCPI command:

SYST:COMM:APR:LEV 4

4.

Set the input source as Local. Send the SCPI command:

SYST:REM:SOUR LOC

5.

Set the main output voltage to 100% by sending the

command:

SOUR:VOLT <ModelVolt>

or use the rotary adjust/Enter control.

6.

Turn the main output On by pressing the OUTPUT ENABLE

Main button on the front panel or send the command:

:OUTP ON

The OUTPUT ENABLE Main button will illuminate.

7.

Enter interactive calibration mode by sending the SCPI

command:

CAL:MON

8.

Press on the terminal keys - or + to reduce or increase

monitoring value. Adjust the monitoring value displayed on
the voltmeter that was set up in Step 1 to read as close as
possible to 4.000 V.

9.

Press any key other than the + or - keys to exit from

interactive calibration mode.