Over voltage protection calibration – AMETEK XG Family Programmable DC User Manual

Page 273

Advertising
background image



Calibration and

Troubleshooting

M370430-01 Rev E

7-9

7

Over Voltage Protection Calibration

Important:

The Voltage Calibration must be done before performing this

procedure. If this is not done the OVP calibration will be inaccurate.

To calibrate the over voltage protection:

1.

Turn on the power supply.

2.

Select the VAP mode to be Off.

3.

Set the voltage set point to be 90% of the model voltage and

the current to be 10% of the model current.

4.

Turn on the main power output.

5.

Execute the OVP calibration SCPI command:

:CAL:PROT

6.

The power supply will enter into self-calibration mode. The

output voltage display blinks OVP CAL. After several
seconds, the calibration will complete and the display will
return to normal operation with output off.

Calibration of the power supply is complete.

Important:

The output (main) voltage must be precalibrated.

The SCPI command (s) for these instructions are:

[:]CALibration[:VOLTage]:PROTection