Auxiliary Auto Start Mode

The Auxiliary Auto Start mode determines the state of the auxiliary
output after a complete power cycle (all front panel LEDS are not
illuminated). With Auxiliary Auto Start mode turned to On, the
auxiliary output will be activated after the power supply is powered up
again.

To define the Auxiliary Auto Start mode:

1.

Press and hold the OUTPUT ENABLE Aux button for 1
second.

or

is displayed.

2.

To change to status to On or Off, press the OUTPUT
ENABLE Aux button.

The SCPI command (s) for these instructions are:

[:]OUTPut[<channel>]:AUXiliary:PowerON[:STATe]

Constant Power Mode

(version 2.21 and higher)

In

Constant Power Mode the power supply will continually adjust the

current levels to maintain a constant power set point. This mode is
intended primarily for loads with response time greater than
approximately 100ms. In order to provide additional protection for the
load, voltage and current limits may be set while in the constant power
mode. If the unit cannot regulate to the Constant Power setting due to
load conditions, it will regulate at either the voltage or the current limit
depending on the load demand.

Example, CP Mode:

Constant Power set 500W

Voltage limit 61V

Current limit 20A

To activate Constant Power mode and set the value:

3.

Turn the 9-position mode control to the FLD position or press
the rotary Adjust/Enter control if the 9-position mode control
is already at the FLD position.

4.

Turn the rotary Adjust/Enter control until “CP” is displayed.

5.

Press the rotary Adjust/Enter control.

6.

Turn the Adjust/Enter control to the desired set point.

7.

Press the rotary Adjust/Enter control

8.

Turn the Adjust/Enter control control for fine adjustment.