AMETEK XG Family Programmable DC User Manual

Page 279

Advertising
background image



Calibration and

Troubleshooting

M370430-01 Rev E

7-15

7

Non-isolated Resistive Programming of Current Calibration

Important:

For maximum accuracy at a specific APG level, follow the

calibration procedure again but use the intended operating resistance level in
steps 2 and 3.

To calibrate the non-isolated resistive programming of current:

1.

Short the main output with a shunt. Connect a multimeter to
measure the voltage across the shunt.

2.

Connect the 4.000 kΩ resistor to the non-isolated connector

(APG -J1), one terminal to resistive programming of current
(J1 .12) and voltage programming of current (J1.10), and the
other terminal to the common (J1.7).

3.

Set the resistive APG level to 4 kΩ The SCPI command is:

SYST:COMM:APR:LEV:CURR 4

4.

Set the input source as Resistive Non-Isolated for Current

Programming. The SCPI command is:

SYST:REM:SOUR:CURR ARES

5.

Turn the main output to On by pressing the OUTPUT

ENABLE Main button on the front panel or send the
command:

:OUTP ON

The OUTPUT ENABLE Main button will illuminate.

6.

Enter the interactive calibration mode by sending the SCPI

command:

CAL:OUTP:ANAL

:CURR:RES

7.

Press the terminal keys - or + to reduce or increase the main

output current value. Adjust the main output current to 100%.

8.

Press any key other than the + or - keys to exit from

interactive calibration mode.

Advertising
This manual is related to the following products: