Table A-9

System Commands (Continued)

Query the Error
Queue for the
number of
entries.

[:]SYSTem[<channel>]:ERRor:COUNt?

Returns the number of errors
currently in the error queue.

Yes

Select Voltage
Analog
Programming
Input Voltage
Input Level/
Range

[:]SYSTem[<channel>]:COMMunicate :APRo

gram:LEVel[:VOLTage][:ISOLated]

Select level for Voltage
APG. Level can be between
2 -10 volts.

?|<Level (2–
10)>|MAXimum|MI
Nimum

Yes

Select Current
Analog
Programming
Input Voltage
Input Level/
Range

[:]SYSTem[<channel>]:COMMunicate :APRo

gram:LEVel:CURRent[:ISOLated]

Select level for Current
APG. Level can be between
2 -10 volts.

?|<Level(2-
10)>|MAXimum|MI
Nimum

Yes

Select the
address to
communicate
with.

[:]SYSTem[<channel>]:COMMunicate
[:MCHannel]:ADDRess

Select the current device to
receive the commands. The
parameter for this command
is the address set on the front
panel under the remote

Yes

Setup the
Voltage control
APG source

[:]SYSTem[<channel>]:REMote:SOURce
[:VOLTage]

Alter or query the state of
the remote programming
voltage APG source.

?|LOCal|AVOLtag

e|IAVoltage|ARE
Sistive|IAResis
tive

Yes