Calibration

and Troubleshooting

7-12

M370430-01 Rev E

Non-isolated Voltage Programming of Voltage Calibration

Important:

For maximum accuracy at a specific APG level, follow the

calibration procedure again but use the intended operating voltage level in steps
3 and 4.

To calibrate the non-isolated voltage programming of voltage:

1.

Disconnect the load and connect the voltmeter to the output.

2.

Connect the input voltage source to the non-isolated

connector, voltage programming (J1.9 - J1.7).

3.

Set the Voltage APG level to 4 V. Send the SCPI command:

SYST:COMM:APR:LEV 4

4.

Set input source voltage to 4.000 V.

5.

Set the input source as Voltage Non-Isolated for Voltage

Programming. The SCPI command is:

SYST:REM:SOUR AVOL

6.

Turn the main output to On by pressing the OUTPUT

ENABLE Main button on the front panel or send the
command:

:OUTP ON

The OUTPUT ENABLE Main button will illuminate.

7.

Enter the interactive calibration mode by sending the SCPI

command:

CAL:OUTP:ANAL

8.

Press the terminal keys - or + to reduce or increase the main

output voltage value. Adjust the main output voltage to 100%,
measured on the external voltmeter on the output.

9.

Press any key other than the + or - keys to exit from

interactive calibration mode.