Step 1: gain calibration, Step 2: offset calibration – AMETEK XG Family Programmable DC User Manual

Page 268

Advertising
background image



Calibration

and Troubleshooting

7-4

M370430-01 Rev E

Step 1: Gain Calibration

M

e

a

s

u

re

d

V

a

lu

e

Programmed Value

Programmed Value

Figure 7-2

Calibration: Step 1 Gain Calibration

Adjust the gain so that the real line and ideal line intersect at a
programmed value of 90%.

Step 2: Offset Calibration

M

e

a

s

u

re

d

V

a

lu

e

Programmed Value

Programmed Value

Figure 7-3

Calibration: Step 2 Offset Calibration

Adjust the offset so that the real and ideal lines intersect at a
programmed value of 10%.