Super Systems SuperDATA User Manual

Page 29

Advertising
background image

Super Systems Inc

Page 29 of 172

Super Data Operations Manual

Note: Items enclosed in brackets [ ] are optional but may be required for certain instrument types.


Item

Definition

C

Optional. Applies to ALL instruments. Specifies the Connection Identifier (1-8
for serial connections and 9 for Ethernet). The protocol for the connection
must match the protocol for the instrument. If not included, defaults to
Connection 1.

LLL

Required. Applies to ALL instruments.

The logical channel number assigned (1-

128).

PPP

Required. Applies to ALL instruments. (see MMI Instrument Physical
Addresses )

ID

Required. Applies to ALL instruments.

The instrument ID: 10PRO or

BARBER-COLEMAN-560

NOLOG

Optional. Applies to ALL instruments.

Do not log data from this channel.

Include this when communications are required for real-time data but are not
required for historical data logging. If this is not included, the channel will be
data logged at one minute intervals.

ABC

Optional. Applies to ALL instruments.

Alarm Block Configuration .

Caution:

any specified channel can use either ALM or ABC but not both.

Allows a custom

Alarm Block Bitmap (one word) to be setup for any instrument but generally
used only for non-programmable instruments that are not capable of
maintaining alarm bitmaps. See the section on Custom Alarm Block
Configuration.

TAG

Required. Applies to ALL instruments.

Instrument name. Limited to 12

Characters with no spaces.




MMI V3 and V3.5 Instruments


[C]CH#LLL(PPP ) = "ID,[PROG],[BC CONT ],[AEVT ],[EVT ],[NOLOG],[ABC] " = "tag "

Examples:

2CH#22(2)=" UCARB-V3.0,PROG(3),BC_CONT,EVT(4)"="FCE_2"

2CH#23(3)=" UNIPRO,PROG(3),EVT(4)"="TEMPER_1"

Note: Items enclosed in brackets [ ] are optional but may be required for certain instrument types.

Item

Definition

C

Optional. Applies to ALL instruments. Specifies the Connection Identifier (1-
8 for serial connections and 9 for Ethernet). The protocol for the connection
must match the protocol for the instrument. If not included, defaults to
Connection 1.

LLL

Required. Applies to ALL instruments.

The logical channel number assigned

(1-128).

PPP

Required. Applies to ALL instruments.

(see MMI Instrument Physical

Addresses )

ID

Required. Applies to ALL instruments.

The instrument ID: MCARB-V3.0,

UCARB-V3.0,

UCARB-V3.5,

MCARB-IR,

MCARB-IR-V3.5,

UNIPRO,

UNIPRO-V3.5

PROG

Optional. Always set this to PROG(3) for V3 and V3.5 Instruments..

BC CONT

Optional. Used to indicate if a V3.x Carb instrument has an attached BC560
or 10Pro controller for temperature.

AEVT

Optional. Used to indicate if a V3.x instrument has attached analog events
on the events port. Indicate the number of analog modules: e.g. AEVT(4).

EVT

Optional. Used to indicate if a V3.x instrument has attached digital events
on the events port. Indicate the number of digital modules: e.g. EVT(4).

Advertising