Parameter power supply (spo2, etco2, nibp), Digital system board, Parameter power supply (spo2, etco2, nibp) -2 – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 154: Digital system board -2, Parameter power supply (spo, Etco, Nibp)

R Series Service Manual
5-2
9650-0903-01 Rev. J
Parameter Power Supply (SpO
2
, EtCO
2
, NIBP)
The parameter power supply board provides power and communications for the Masimo SpO
2
module, the EtCO
2
Capnostat 5 sensor, and the Suntech
Advantage A+ NIBP module. SpO
2
and EtCO
2
power and communications are isolated from the system to meet the requirements for applied parts in
EN60601-1, EN60601-2-4 and EN60601-2-49. NIBP isolation requirements are met by the applied parts themselves; the cuffs and tubing are non-
conductive.
A barometric pressure sensor is included for EtCO
2
monitoring. In addition, a hardware R wave trigger is routed from the digital system board to the NIBP
module.
Digital System Board
The R Series Digital System Board is the main control unit of the R Series defibrillator. The digital system includes these major blocks:
•
System Processor: SH3-DSP SH7229
•
FLASH memory for program boot
•
SDRAM for program and data storage
•
USB Device and Host interfaces
•
One compact flash interface
•
Disk-On-Chip for non-removable data storage
•
Display Controller: 8M RAM, LCD and VGA outputs
•
CPLD for system control
•
FPGA for I/O control and serial I/O
•
CODEC for audio input (microphone: future) and output (speaker)
•
ECG Out
•
RS232 (internal use only)
•
Beeper driver
•
Strip Chart Printer controller and drivers
•
Real time clock / calendar with precision crystal and battery
•
Safety functions including watchdog, power monitors and reset control.
The digital system board controls and/or communicates with external functional modules and real time peripherals including the core P/D Engine, core
ECG, parameters (SpO
2
, EtCO
2
and NIBP), CPR, AC charger, battery, and controls board.