Parameter power supply (spo2, etco2, nibp), Digital system board, Parameter power supply (spo2, etco2, nibp) -2 – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 154: Digital system board -2, Parameter power supply (spo, Etco, Nibp)

Advertising
background image

R Series Service Manual

5-2

9650-0903-01 Rev. J

Parameter Power Supply (SpO

2

, EtCO

2

, NIBP)

The parameter power supply board provides power and communications for the Masimo SpO

2

module, the EtCO

2

Capnostat 5 sensor, and the Suntech

Advantage A+ NIBP module. SpO

2

and EtCO

2

power and communications are isolated from the system to meet the requirements for applied parts in

EN60601-1, EN60601-2-4 and EN60601-2-49. NIBP isolation requirements are met by the applied parts themselves; the cuffs and tubing are non-
conductive.

A barometric pressure sensor is included for EtCO

2

monitoring. In addition, a hardware R wave trigger is routed from the digital system board to the NIBP

module.

Digital System Board

The R Series Digital System Board is the main control unit of the R Series defibrillator. The digital system includes these major blocks:

System Processor: SH3-DSP SH7229

FLASH memory for program boot

SDRAM for program and data storage

USB Device and Host interfaces

One compact flash interface

Disk-On-Chip for non-removable data storage

Display Controller: 8M RAM, LCD and VGA outputs

CPLD for system control

FPGA for I/O control and serial I/O

CODEC for audio input (microphone: future) and output (speaker)

ECG Out

RS232 (internal use only)

Beeper driver

Strip Chart Printer controller and drivers

Real time clock / calendar with precision crystal and battery

Safety functions including watchdog, power monitors and reset control.

The digital system board controls and/or communicates with external functional modules and real time peripherals including the core P/D Engine, core
ECG, parameters (SpO

2

, EtCO

2

and NIBP), CPR, AC charger, battery, and controls board.

Advertising