Operator safety – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 189

Advertising
background image

Warnings

9650-0904-01 Rev. K

ZOLL R Series Operator’s Guide

1–17

Do not use the unit in AED mode or advisory mode during patient movement. A patient must
be motionless during ECG rhythm analysis. Do not touch the patient during analysis. If
transporting the patient, cease all movement before beginning ECG analysis.

ECG rhythm analysis does not warn of patient asystole, which is not a shockable rhythm.

The ECG rhythm analysis function might not reliably identify ventricular fibrillation in the
presence of an implanted pacemaker. Inspection of the electrocardiogram and clinical evidence
of cardiopulmonary arrest should be the basis for any treatment of patients with an implanted
pacemaker.

Implanted pacemakers might cause the heart rate meter to count the pacemaker rate during
incidents of cardiac arrest or other arrhythmias. Dedicated pacemaker detection circuitry may
not detect all implanted pacemaker spikes. Check the patient's pulse; do not rely solely on heart
rate meters. Patient history and physical examination are important factors in determining the
presence of an implanted pacemaker. Pacemaker patients should be carefully observed.

Do not place electrodes directly over an implanted pacemaker.

The R Series unit detects ECG electrical signals only. It does not detect a pulse (effective
circulatory perfusion). Always verify pulse and heart rate by physical assessment of the patient.
Never assume that the display of a nonzero heart rate means that the patient has a pulse.

To avoid possible damage to the R Series unit, turn off pacing before defibrillating the patient
with a second defibrillator.

Do not use the unit’s ECG-out signal as a synchronization pulse for another defibrillator or
cardioverter.

Place the patient on a firm surface before performing CPR.

Battery

Although the device can operate with AC power alone, ZOLL strongly recommends that you
operate the unit with a battery installed at all times. Operating the unit with a battery provides a
backup in case of AC power shortage, and results in faster charge time. The battery can be
automatically recharged while it is installed in the unit. Keep a fully charged spare battery pack
with the defibrillator at all times.

Test battery packs regularly. A battery that does not pass the ZOLL charger’s capacity test
might cause the R Series unit to shut down unexpectedly.

When the warning LOW BATTERY appears, plug the R Series unit into a power source or install
a fully charged battery pack. When the warning REPLACE BATTERY appears, immediately
replace the battery pack with a fully charged pack or plug the R Series unit into a power source,
as unit shut down due to a low battery condition is imminent.

If mistreated, a battery pack might explode. Do not disassemble a battery pack or dispose of it
in fire.

Operator Safety

Do not use R Series products in the presence of oxygen-rich atmospheres, flammable
anesthetics, or other flammable agents (such as gasoline). Using the unit in such environments
might cause an explosion.

Do not use the unit near or within standing water. Electrical safety might be compromised when
the defibrillator is wet.

Advertising