Recorder check, Code readiness log – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 488

Advertising
background image

C

HAPTER

12

M

AINTENANCE

12–6

www.zoll.com

9650-0912-01 Rev. K

The messages CHECK PADS and POOR PAD CONTACT appear alternately. The pace alarm
sounds, and the Clear Pace Alarm softkey flashes.

6.

Reconnect the OneStep cable, and press the Clear Pace Alarm softkey.

The messages CHECK PADS and POOR PAD CONTACT disappear, and the alarm tone
stops.

Recorder Check

1. Check the printer for an adequate supply of paper, then press the RECORDER button.

2.

Press and hold the SIZE button for at least 2 seconds.

A calibration pulse of 1 mV appears on the display while the button is held. The amplitude
of the calibration pulse is independent of the SIZE setting.

3.

Inspect the recorder waveform for uniformity and darkness.

4.

Inspect for uniformity of annotated characters and completeness of words.

5.

Check the printer speed by verifying that the calibration pulse is:

2.5 mm ±0.5 mm wide

10 mm ±1.0 mm high

Code Readiness Log

Each automatic and manual defibrillation test result is stored in internal non-volatile memory,
the Code Log. A total of 1000 Code Readiness test records can be stored in internal memory.
When the Code Readiness Log is full, the oldest records are erased on a first-in-first-out (FIFO)
basis.

The Code Readiness Log can be transferred to an external computing device (see “Transferring
Files to an External Device” on page 11-1).

If configured, the R Series prints a Code Readiness Test Report following the completion of
each automatic defibrillator test.

To print the Code Readiness Log:

1. In MONITOR mode, press the Report Data softkey, then the Test Log softkey.

A menu appears with the print options.

2.

To print a specific test, use the Prev Test and Next Test softkeys to select the test, then press
Print Test, or press Print Test Log to print the log of all of the tests.