0 synchronized cardioversion test, Synchronized cardioversion test -23 – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 33

Advertising
background image

R Series Service Manual

9650-0903-01 Rev. J

1-

23

12.0 Synchronized Cardioversion Test

Tools Needed: Impulse 4000 or equivalent defibrillator analyzer.

Test Setup:

1. Connect the OneStep cable via the adapter (D.N.I #3010-0378) to the defibrillator analyzer.

2.

Select cardioversion on analyzer. Input 1mV ECG signal at 60 BPM.

Do this...

Observe this...

Pass/Fail

12.1

Press LEAD button to select PADS and Size
X1.

12.2

Press the SYNC softkey on the defibrillator.
Enter synchronized cardioversion timing test
mode on the defibrillator analyzer.
NOTE: Press the SYNC softkey again if the unit
is configured for Remote Sync.

Sync appears on display.

Sync markers display on the monitor. The sync marker appears as a down
arrow over the ECG R-wave peaks on strip chart and display.

o

o

12.3

Select 200J.

o

o

12.4

Press the CHARGE button. When the SHOCK
button lights, press and hold the SHOCK
button.

Defibrillator discharges.

Observe that the R-wave to shock delay (sync delay) is less than 60
milliseconds on the analyzer display.

o

o

Record your results on the Maintenance Test Checklist.