Defib 120j sel, Shock advised, 3 press shock – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 421

Advisory Defibrillation Procedure
9650-0912-01 Rev. K
ZOLL R Series Operator’s Guide
4–5
•
Units with the automatic charge option disabled will alternately display the messages
SHOCK ADVISED and PRESS CHARGE. Press the CHARGE button.
Regardless of the analysis result, the user can control the defibrillator manually. For example,
the user can defibrillate the patient even if the advisory function issues a NO SHOCK ADV.
message.
3 Press SHOCK
WARNING!
Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.
Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient’s
body to come in contact with metal objects, such as a bed frame, as unwanted pathways
for defibrillation current may result.
DEFIB 120J SEL.
Code
Marker
Param
Sync
On/Off
00:01
SHOCK ADVISED
ECG
Options
Alarms
Report
Data