ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 36

Advertising
background image

R Series Service Manual

1-26

9650-0903-01 Rev. J

14.11

Press the ENERGY SELECT up arrow until
200J displays.

Verify that DEFIB 200J SEL is displayed.

o

o

o

14.12

Press the CHARGE button. Wait for the
SHOCK button to illuminate.

Verify that DEFIB 200J RDY is displayed.

o

o

o

14.13

Press the SHOCK button.

Verify the unit discharges 196J-264J into the simulator.

o

o

o

14.14

Press the CHARGE button and start timing with
a stopwatch. Stop timing when the SHOCK
button illuminates.

Observe and record the value of the charge time on the stop watch.

Charge time 1.0-7.0 sec.

o

o

o

14.15

Press the SHOCK button.

Verify that the Patient Current is between 23.9-25.9A and Defib Impedance is
between 46-54 Ohms on the strip chart.

o

o

o

14.16

Press the CHARGE button, when the SHOCK
button Illuminates, press the ENERGY SELECT
DOWN ARROW
.

Verify that no fault messages are displayed, and no discharge energy is
displayed on the Impulse 4000.

o

o

o

14.17

Set Energy level to 200 Joules. Press the
CHARGE button. When the SHOCK button
illuminates, start timer.

Verify the unit holds the energy for 60 seconds (15 seconds for AED units), and
that when the unit discharges internally, that there are no fault messages
displayed.

o

o

o

14.18

Disconnect the cable from the analyzer.

CHECK PADS audio prompt.

o

o

o

Record your results on the Maintenance Test Checklist.

Do this...

Observe this...

Pass/Fail/NA

Advertising