ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 64

Advertising
background image

R Series Service Manual

2-12

9650-0903-01 Rev. J

SHOCK ADVISED

Advisory message when Analysis finds a shockable
rhythm. Followed by PRESS SHOCK in Manual Advisory
Defib with auto-charge enabled or in Auto Defib mode, or
by PRESS CHARGE in Manual Advisory Defib with auto-
charge disabled.

SPO2 AMBIENT LIGHT

Ambient light is too bright.

Shield sensor from ambient light.

Replace sensor.

Replace SpO

2

module

SPO2 COMM ERROR

No transmissions from SpO

2

unit received.

Communication error or no communication from SpO

2

module.

Replace SpO

2

module and/or system board.

SPO2 PULSE SEARCH

Pulse search in progress.

STAND CLEAR

(Manual Advisory Defib with auto-charge enabled or Semi-
Auto Mode Defib) User pressed Analyze or an analysis
was started automatically as part of the rescue protocol.
Patient rhythm is being analyzed.

SUMMARY TIMEOUT

Summary report timeout error.

SYNC DEFIB DISABLED

Sync mode active when analyze pressed in defib.

SYSTEM FAULT 210

System task have not been activated for 500 ms – ECG
control.

Replace digital board.

SYSTEM FAULT 211

System task have not been activated for 500 ms – Defib.

Replace digital board.

SYSTEM FAULT 212

System task have not been activated for 500 ms – Pace.

Replace digital board.

SYSTEM FAULT 212

System task have not been activated for 500 ms – User
Interface.

Replace digital board.

SYSTEM FAULT 214

System task have not been activated for 500 ms – Display.

Replace digital board.

SYSTEM FAULT 215

System task have not been activated for 500 ms – Not
used.

Replace digital board.

Error Message

Explanation

User
Advisory

Technical Action