ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 63

R Series Service Manual
9650-0903-01 Rev. J
2-11
REPEAT NIBP MEAS
•
The unit exceeded the maximum number of inflation
attempts.
•
The unit exceeded the 180-second measurement time
limit.
•
Check cuff and hose.
•
Repeat NIBP measurement.
REPLACE BATTERY
Battery voltage is less than absolute minimum. Shutdown
imminent.
Replace with charged battery.
REPORT FULL
Summary report memory full.
Erase summary report.
RETRY ANALYSIS
Advisory message in conjunction with noisy ECG. Analysis
halted.
RPT NO DISK
Internal disk is not formatted and cannot be formatted.
SELECT 30J FOR TEST
Attempt to run a self test at an energy other than 30J.
SELECT ASYNC PACE
User is prompted to select Async pacing, since the PD
module cannot detect sync pulses.
SELECT DEFIB MODE
Analyze button pressed in pace or monitor mode.
Turn main selector knob to Defib/ON.
SELECT LIMB LEADS
Paddles or augmented ECG leads selected when
continuous analysis active or started.
Select limb leads I, II, III or MFE.
SELECT PADS
ECG lead (non-MFE Pads) selected when ANALYZE
pressed.
SET CLOCK
Real time clock failure: invalid date or time.
•
Set time and date information.
•
Verify that the internal lithium battery has been
replaced within the last 5 years. Contact ZOLL
Technical Service Department for assistance.
SET PACE MA
Multiple copy errors are the product of intended software
or memory errors. If error reoccurs other than on entering
pace the first time or after more than 10 minutes in other
mode, the unit could be broken.
Set pace current. If broken, replace system board.
Error Message
Explanation
User
Advisory
Technical Action