Defibrillator specifications – ZOLL R Series Monitor Defibrillator Rev J User Manual

Page 328

Advertising
background image

A

PPENDIX

A

S

PECIFICATIONS

A–2

www.zoll.com

9650-0904-01 Rev. K

Defibrillator Specifications

General

Size
(height • width • length)

8.2 in. • 10.5 in. • 12.5 in. with handle or 10.0 in. without handle
20.8 cm • 26.7 cm • 31.7 cm with handle or 25.4 cm without handle

Weight

13.6 lb (6.17 kg) with OneStep cable and battery pack
15.2 lb (6.89 kg) with paddles

Power (for R Series ALS,
BLS and Plus models)

Battery: Rechargeable lithium ion battery pack

AC Power Requirements: 100-120 Vrms, 50/60 Hz

220-240 Vrms, 50 Hz

AC Power Consumption: 275 VA maximum

Device classification

Class I and internally powered per EN 60601-1.

Design standards

Meets or exceeds applicable requirements of UL 60601, AAMI DF80,
IEC 60601-2-4, EN 60601-2-25, and EN 60601-2-27.

Patient safety

All patient connections are electrically isolated.

Environmental

Operating temperature

0°C to 40°C (32°F to 104°F)

Storage and shipping
temperature

–20°C to 60°C (-4°F to 140°F)

Humidity

5% to 95% relative humidity, noncondensing

Vibration

IEC 68-2-6 and IEC 68-2-34

Shock

IEC 68-2-27, 50 g 6ms half sine

Operating pressure

594 to 1060 millibars (-1253 to 14046 ft.)

Particle and water
ingress

IEC 529, IP 22

Electromagnetic
compatibility (EMC)

CISPR 11 Class B - radiated and conducted emissions

Electromagnetic
immunity

AAMI DF80; EN 61000-4-3 to 10 V/m

Electrostatic discharge

AAMI DF80; EN 61000-4-2

Conducted susceptibility

EN 61000-4-4, 61000-4-5, 61000-4-6

Display

Screen type

High-resolution, liquid crystal display (LCD)

Screen size

6.5 inches (16.5 cm) diagonally

Display format

Nonfade moving bar display.

Sweep speed

25 mm/s

Viewing time

5 seconds
4 seconds with certain monitoring parameter options